Electroencephalography for the Sleep Specialist

Chapter 2


Electroencephalography for the Sleep Specialist



An understanding of the technical aspects of electroencephalography (EEG) and recognition of basic normal and abnormal EEG patterns in wakefulness and sleep are essential skills for a polysomnographer. EEG records the difference in electrical potentials between two electrodes placed over the scalp. This scalp EEG activity results from extracerebral current flow caused by summated excitatory postsynaptic potentials (EPSPs) and inhibitory postsynaptic potentials (IPSPs). Rhythmical oscillations of the thalamocortical neuronal projections cause synchronous synaptic EPSPs and IPSPs over areas of the cortex. The scalp EEG recording voltages are attenuated by the skull and intervening tissues and reflect about one tenth of the voltage recorded over the cortical surface.



Method of Electroencephalographic Recording


Electrical signals are transmitted by the electrodes and conducting gel through electrode wires, which connect the electrodes and the jack box of the polysomnography (PSG) equipment. Different types of electrodes are available, but for recording EEG the best electrodes to use are gold cup electrodes with holes in the center and the silver–silver chloride electrodes. Silver–silver chloride electrodes need repeated chloriding for proper maintenance. Electrode paste or conducting gel is used to secure the electrodes. Positive and negative charges are generated between the scalp and recording electrode as a result of ionic dissociation. The electrode-electrolyte interface is the most critical link in the PSG machine, because most artifacts originate at this site; careful preparation is therefore very important. The impedance in a pair of electrodes should be measured by an impedance meter and should not be greater than 5000 ohms. High impedance impairs the ability of the electrical signal to reach the amplifier and interferes with the capacity of the amplifier to eliminate environmental noise, thus increasing artifacts. All electrode wires terminate in a pin, called a jack, that is plugged into the electrode box. All the wires from the head are gathered into a ponytail in back of the head and then secured by tying them together. A wrap around the scalp secures the electrodes and the wires. The arrangement is therefore as follows: electrodes from the surface of the scalp are connected by electrode wires into the jacks, which are numbered, and then the wires from the jack box are connected through a shielded conductor cable to the electrode montage selector containing rows of switches in pairs corresponding to the inputs of the amplifier.


The placement of the electrodes is determined by the 10-20 electrode placement system that is recommended by the International Federation of Societies for EEG and Clinical Neurophysiology and was published by Jasper in 1958. The 10-20 system is based on definable anatomical landmarks (Fig. 3.1). The system consists of letters denoting the parts of the brain underneath the area of the scalp and numbers denoting specific locations. The odd numbers refer to the left side of the head, and the even numbers refer to the right side. The technician should be familiar with the measurement technique for placement of the electrodes according to the 10-20 international system. Important landmarks for measuring include the inion, the nasion, and the right and left preauricular points. The distance from the nasion to inion along the midline through the vertex should be measured. FPz in the midline is 10% above the nasion of the total distance between the inion and nasion. The electrodes marked FP1 and FP2 are located laterally 10% above the nasion of the total distance between the inion and nasion measured along the temporal regions through the preauricular points. Oz denotes an electrode placed at a distance of 10% above the inion of the total distance between the nasion and inion. T3 and T4 electrodes are placed in a location 10% above the preauricular points of the total distance between the two preauricular points. The rest of the electrodes are located at a distance of 20% measured from inion to nasion anteroposteriorly or laterally through the ears as well as transversely between the ears. The montage refers to an arrangement of two electrodes (derivations). Both bipolar montage (connection of the electrodes between two relatively active sites over the scalp) and referential montage (connection of the electrodes between an active and relatively inactive site, e.g., M1, M2, Cz, Pz; see Fig. 2.1) are recommended. The nomenclature was recently changed to rename T3, T4, T5, and T6 to T7, T8, P7, and P8, respectively, in a modified 10-20 electrode placement system.



The 2007 American Academy of Sleep Medicine manual recommends three channels; this results in significant limitations (see Chapter 1). Most importantly, such a limited montage would result in the inability to capture focal epileptiform activity arising from the temporal lobes, the most common location for such discharges, as well as missing focal slowing from the hemisphere not being recorded. In our own laboratory we use between four and eight EEG channels, including temporal leads from both hemispheres (see Table 1.2); this increases the yield of capturing focal or diffuse slow waves or epileptiform activities over recording with two to four channels. Many patients are referred to the sleep laboratory with a diagnosis of possible nocturnal seizures. In these cases an extended EEG montage (Fig. 2.1; see Table 1.3), covering the bilateral temporal and parasagittal regions and including both bipolar and referential channels, is recommended. A full complement of electrodes and special electrode placements (e.g., Tl and T2 electrodes) should be used. Simultaneous video recording (video-PSG study) for correlation of EEG activities with the actual behavior of the patient is crucial. In computerized PSG recordings (digital PSG recordings), which are currently performed in most laboratories, it is easy to change the recording speed from the standard 10 mm/sec of the usual sleep recording to 30 mm/sec of the standard EEG recording for proper identification of the epileptiform discharges.



Normal Waking and Sleep Electroencephalographic Rhythms in Adults


The dominant rhythm in adults during wakefulness is the alpha rhythm, consisting of 8 to 13 Hz activity distributed synchronously and symmetrically over the parietooccipital regions (see Fig. 2.1). The frequency of the rhythm between the two hemispheres should not vary by more than 1 Hz, and the amplitude should not vary by more than 50%. This alpha rhythm is best seen during quiet wakefulness with eyes closed and is significantly attenuated by eye opening or mental concentration. A small percentage of normal adults show no alpha rhythm, and their EEG is characterized by a dominant rhythm of low-amplitude EEG in the beta frequencies (greater than 13 Hz) with amplitudes varying between 10 and 25 μV (Fig. 2.2). In most adults the beta rhythms are seen predominately in the frontal and central regions intermixed with the posterior alpha rhythms.



There are characteristic changes in the background EEG rhythms as an individual progresses through the three stages of non–rapid eye movement (NREM) sleep and the tonic and phasic stages of rapid eye movement (REM) sleep. The various characteristics during different sleep stages are described in detail in Chapter 3.


EEG in older adult subjects shows a progressive slowing of the alpha frequency during wakefulness and diminution of alpha blocking and photic driving responses. Focal temporal slow waves, particularly in the left temporal region, often called benign temporal delta transients (Fig. 2.3) of older adults, are seen in many apparently normal older adults, and are sometimes associated with sharp transients. Transient bursts of anteriorly dominant rhythmical delta waves may also be seen in some older adults in the early stage of sleep. Other changes in older adults consist of sleep fragmentation with frequent awakenings, including early morning awakening and multiple sleep-stage shifts. Another important finding in the sleep EEG of older adults is the reduction in amplitude of the slow waves. Because of this, many slow waves do not meet American Academy of Sleep Medicine scoring guidelines; the percentage of slow-wave sleep is therefore often reduced in these subjects.




Abnormal Electroencephalographic Patterns


Abnormal EEG patterns consist of focal or diffuse slow waves and epileptiform discharges. The importance of multiple-channel EEG recordings (see Table 1.3) is to document this focal (Fig. 2.4) or diffuse (Figs. 2.5 and 2.6) slowing and epileptiform discharges (see later).







Electroencephalographic Signs of Epilepsy


EEG is the single most important diagnostic laboratory test for patients with suspected seizure disorders. Certain characteristic EEG waveforms correlate with a high percentage of patients with clinical seizures and therefore can be considered of potentially epileptogenic significance. These epileptiform patterns consist of spikes, sharp waves, spike and waves, sharp and slow-wave complexes, as well as evolving pattern of rhythmical focal activities, particularly in neonatal seizures. In addition, a pattern that correlates highly with complex partial seizure is temporal intermittent rhythmical delta activity (TIRDA) (Fig. 2.7). Another pattern that is considered a marker of the seizure-onset zone is interictal scalp high-frequency oscillations (HFOs) consisting of gamma frequency activity (30-80 Hz), ripples (80-250 Hz), and fast ripples (250-1000 Hz), recorded noninvasively using amplifiers with appropriate filter settings (Fig. 2.8). Therefore recording HFOs may help map epileptogenic zones.



< div class='tao-gold-member'>

Stay updated, free articles. Join our Telegram channel

Jul 16, 2016 | Posted by in NEUROLOGY | Comments Off on Electroencephalography for the Sleep Specialist

Full access? Get Clinical Tree

Get Clinical Tree app for offline access