Diameters (mm)
2.75, 3.0, 3.5, 4.0, 4.5, 5.0
Lengths (mm)
12, 16, 19, 26
Maximum guidewire
0.014″
Shaft size
Proximal: 3.1 Fr
Distal: 2.6 Fr
Minimum deployment pressure
14 atm
Rated burst pressure
16 atm
Vascular Access
Vascular access is gained as described above using a 21G entry needle and 4-Fr micropuncture set.
A 6-Fr or 7-Fr Guide catheter (e.g., Envoy) connected to a continuously running flush of heparinized saline, may be used instead of the shuttle sheath.
Advance the Guide catheter over a 0.035 guidewire (e.g., 0.035 glidewire) through a short sheath using fluoroscopy and roadmapping.
Position the distal tip of the guide catheter proximal to the lesion and remove any slack.
Ensure that the tip of guide catheter is close enough to the lesion to enable stable stent deployment. However, there should be enough space between the lesion and guide tip to enable unobstructed deployment.
Remove the system carefully from its packaging maintaining sterile precautions.
Remove the delivery system out of the sterile hoop ensuring it is not kinked, or bent in the process.
Hold the delivery system just proximal to the balloon (at the proximal balloon bond site) with one hand, and with the other hand, gently remove the stent protector and stylet located distally.
Intervention
Ex Vivo Preparation of Balloon
Fill a 20- or 30-ml luer lock syringe with diluted contrast medium (2/3:1/3 concentration).
Attach the syringe to inflation device using a three-way stopcock and aspirate the contrast medium into the inflation device leaving 2–3 ml in the syringe.
Ensure that the entire inflation system is free of air bubbles.
Attach the system to the balloon inflation port of the delivery system.
Turn the stopcock so that the inflator is occluded and there is direct communication between syringe and balloon catheter.
Aspirate to remove air from the system, holding the syringe upright. Gently allow the plunger to return toward its original position so that air is replaced by contrast in the stent system.
Repeat the process, until air bubbles are no longer visible during aspiration.
Allow the ambient pressure to return to the balloon before detaching the syringe.
A meniscus will appear in the balloon inflation port when the syringe is detached.
Purge the aspirated air from the inflation system by inverting the syringe and injecting out the air.
If air bubbles continue to be aspirated despite multiple aspirations, discard the system and prepare another.
We usually perform an in vivo preparation, which is described below in the sequence it is undertaken during procedure.
Moisten the stent system by soaking it in a bowl containing heparinized saline (DO NOT wipe down with moist gauze sponge as the stent graft may be disrupted by gauze fibers).
Visually confirm that the stent graft is located between the proximal and distal stent markers.
Stent Delivery
Place a 0.014″ microguidewire into the lumen of the balloon catheter.
Ensure the appropriately positioned guide catheter (distal tip of the guide catheter should be proximal to the lesion) is stable.
Advance this system through the RHV of the guide catheter.
Using fluoroscopy and roadmapping, advance the delivery system to the target lesion and position it optimally such that the distal radiopaque marker on the balloon catheter should be distal to the lesion and the proximal marker, proximal to the lesion.
Ensure that the distal marker is at least 1 mm distal to the lesion.
Once optimal positioning is confirmed sufficiently tighten the RHV around the system. Do not over tighten, which may impede the proper functioning of the delivery system.
If ex vivo balloon preparation was performed, then securely attach the inflation system to the balloon inflation port by making a wet meniscus to meniscus connection. To do this, fill the port with diluted contrast to replace any air, and then make a meniscus to meniscus connection with the inflation device.
Ensure that inadvertent, premature deployment of the balloon catheter does not occur during attachment.
In Vivo Balloon Preparation
If ex vivo preparation was not performed, then do in vivo preparation as follows:
Fill a 20- or 30-ml luer lock syringe with diluted contrast medium (2/3 contrast:1/3 saline).
Attach the syringe to inflation device using a three-way stopcock and aspirate the contrast medium into the inflation device leaving 2–3 ml in the syringe.
Ensure that the entire inflation system is free of air bubbles.
Securely attach the system to the balloon inflation port of the delivery system.
Turn the stopcock so that the inflator is occluded and there is direct communication between syringe and balloon catheter.
Aspirate to remove air from the system, holding the syringe upright. Gently allow the plunger to return toward its original position so that air is replaced by contrast in the stent system.
Repeat the process, until air bubbles are no longer visible during aspiration.
Stent Deployment
Prior to stent deployment, confirm that the stent is appropriately positioned and the guide catheter and balloon catheter are free of slack.
Turn the stopcock so that it is off to the syringe and open to inflation device and balloon catheter port.
Place the compliance chart in front of you.
Turn the inflator slowly (approximately 1 atm per 15 s) to gradually inflate the balloon. Inflate to a minimum of 14 atm to expand the stent by balloon inflation.
Use the compliance chart to achieve the appropriate inflation pressure for a particular stent size.
Do not exceed the rated burst pressure or expand the stent beyond its actual size.
Maintain the inflation pressure for 15–30 s for full expansion and adequate approximation of the stent to vessel wall. Underexpansion may result in stent graft movement or lack of proper coverage of perforation.
After completion of stent deployment, deflate the balloon completely by applying negative pressure, if needed. Wait at least 15 s to ensure complete deflation.
Confirm fluoroscopically that the balloon is completely deflated.
Fully open the RHV of the guide catheter.
Maintaining negative pressure, withdraw the balloon catheter from the deployed stent. Ensure that the guidewire and guide catheter maintain their position.
Close the RHV around the guidewire, once the balloon catheter is removed.
Perform angiography to confirm complete apposition of stent against vessel wall.
If complete apposition was not achieved, then use a non-compliant, high-pressure balloon of the same size (or slightly larger) and length as the balloon of the delivery system to perform a second stent expansion.
Do not overdilate the vessel to the extent that dissection may occur.
The final post-procedure angiogram should demonstrate the stent diameter to be the same as (or slightly larger) than the vessel proximal and distal to the stent.
Observe the patient (especially if procedure done under moderate sedation) and perform periodic angiographic assessment within the initial 30 min after stent deployment.
Perform post-procedure cerebral angiography to rule out any embolic complication.
Do not perform MRI for approximately 8 weeks (until the stent has completely endothelialized), to avoid the risk of stent movement.
Problems Encountered and Solutions
Table 10.2 shows some problems encountered and solutions.
Table 10.2
Some problems encountered and solutions
Problem
Solution
Resistance encountered during advancement of balloon catheter through guide catheter
Do not force your way forward. If possible and already achieved, maintain guidewire access across the lesion and remove the delivery system (balloon catheter and stent) as a single unit. Identify the cause of resistance. If required, replace the system with a new one
A single stent cannot be positioned to cover the lesion adequately
If needed, deploy two stents in a telescopic fashion to cover the lesion entirely. Preferably, deploy the distal stent first, followed by the proximal stent. This will eliminate the need to cross the just deployed proximal stent with a stent delivery system that could result in its dislodgment
Carotid Artery Angioplasty and Stenting
Indications and Case Selection
Carotid Stenosis: If a patient has expected survival of less than 5 years; patients with malignancies and other life-threatening conditions should be considered for endovascular treatment instead of carotid endarterectomy.
Carotid Dissection/Pseudoaneurysm: e.g., due to trauma. In such a case stenting (with or without coiling) may be performed to appose the walls of the injured artery. Angioplasty is not done. Intervention is needed when the dissection is symptomatic; the intimal flap faces the flow of blood, in which case the dissection/associated false lumen is likely to expand; and pseudoaneurysm, will be at a risk of rupture if not treated.
Contraindications
If anticoagulant and/or antiplatelet therapy is contraindicated.
Severe vascular tortuosity or anatomy that would preclude the safe introduction or maintenance of a guide catheter, sheath, or interventional devices.
Hypersensitivity to nickel-titanium.
Uncorrected bleeding disorders.
Lesions in the ostium of CCA.
Preoperative Management
Verify lab values including platelet count, BUN, CR, APTT, PT/INR, and ß-HCG for females of reproductive age group.
In case of renal insufficiency, diabetes, CHF, etc., ensure usage of diluted non-ionic contrast agent and carefully pre-plan to maintain contrast load to minimum.
Plavix 75 mg PO daily starting 3 days before procedure or
Plavix 300 mg (4 tabs) PO LD on morning of procedure or
If unable to administer Plavix in a timely fashion, another option is:
Abciximab (Rheopro®) in a L.D. of 0.25 mg/kg followed by continuous IV infusion of 0.125 µg/kg/min (to a maximum of 10 µg/kg/min) for 12 h, then d/c. Start this after arterial access has been secured. Administer Plavix 75 mg PO the same evening and then 75 mg PO daily.
ASA 81–325 mg PO daily.
Liquids only on morning of procedure.
NPO (for ≈6 h) when procedure performed under GA.
Obtain informed consent for angiography, angioplasty, and stenting.
Ensure two IV lines inserted.
Insert Foley. Patient will be more comfortable and cooperative with an empty bladder in case the procedure becomes prolonged.
Position patient on neuroangiography.
Attach patient to pulse oximetry and ECG leads for monitoring O2 saturation, HR, cardiac rhythm respiratory rate, and BP.
Carotid Stenosis
Drugs
0.9% NS + 20 meq KCl @ 75 cc/h (adjust to higher rate as needed).
Fentanyl 25–100 µg IV prn.
Versed 0.5–1 mg IV prn.
Heparin IV per physician instructions. (See “Technique” below.)
Devices
Micropuncture kit 4 Fr.
6-Fr short sheath (10 cm); use long sheath if tortuous vasculature, e.g., a 6-Fr shuttle sheath which may be advanced initially over dilator into the descending aorta and then into common carotid over 6-Fr H1 slip catheter. A shuttle sheath is also preferable because it will provide greater support and significantly decrease the likelihood of the apparatus collapsing into the aortic arch at critical time points, e.g., during stent deployment.
6-Fr H1 slip catheter.
Rotating hemostatic valves (2). Ensure the RHV attached to the Guide catheter is ≥0.096″ or 2.44 mm.
Pediatric transducers (30 ml/h; 2).
Guide catheter: 6-Fr Envoy® MPC guide catheter (90 cm).
Diagnostic catheter: Terumo® front-angled glidecath 5 Fr (for diagnostics. May use Envoy guide catheter for the same purpose).
Front-angled glidewire (0.035; Terumo).
Embolic protection system (Guidant, Accunet). The package containing this product has three pouches which contain the following: RX Accunet™ delivery system comprising of 0.014″ guidewire with filter basket, peel away delivery sheath, introducer tool, torque device with peel away adapter, and flushing tool.
The other two pouches contain filter basket recovery catheters. One has a shapeable tip, while the other is low profile and more flexible.
Selection of filter basket size will depend upon the diameter of carotid artery. (See chart in “Technique” section.)
Balloon dilatation catheter: Aviator™ Plus (RX) (Cordis, Warren, NJ, USA).
It is a rapid exchange (RX) catheter.
It is available in 4–7 mm balloon diameter range. Select the appropriate size for the vessel being treated, based on the table provided with the product.
Carotid stent: Acculink™ (Guidant). See tables in “Procedure” section for selecting the appropriate size to the vessel being treated, in untapered or tapered stents.
Syringes 10 cc (at least 3), 20 cc (at least 4), 3 cc (for ACT and angioplasty balloon preparation).
Three-way stopcock: 3.
Torque device.
Telfa strip.
Mandrel for shaping microwire tip.
Angioseal™ closure device (6 Fr). Use larger size if a larger sheath inserted.
Vascular Access
Attach a 6-Fr shuttle sheath to a continuously running flush of heparinized saline and confirm that the system is free of air bubbles.
Insert the provided dilator into the shuttle sheath.
Gain access with micropuncture needle using modified Seldinger technique.
Insert a glidewire into the descending aorta.
Remove the 4-Fr sheath over the wire, maintaining wire access to the vasculature.
Introduce the shuttle sheath-dilator assembly onto the glidewire.
Using fluoroscopy, advance the shuttle sheath into the aortic arch, ensuring that the tip of the glidewire is always leading.
Retract and remove the dilator.
Administer 5000 IU heparin IV. Remember to perform ACT approximately 20–30 min later. Using a 15-cc syringe, draw 10–15 cc of blood from the sheath and then immediately attach a 3-cc syringe to draw 1–2 cc of blood, which is sent for ACT.
May return the blood drawn prior to ACT sample back to the patient via the sheath, as long as the distal tip of the sheath is proximal to the carotid vasculature.
Check ACT hourly throughout the procedure and administer heparin IV to maintain an ACT of 300–350.
Attach a 6-Fr H1 slip catheter to a continuously running flush of heparinized saline and confirm that the system is free of air bubbles.
Advance the slip catheter over the guidewire into the sheath if an exchange length wire has been used. Ensure that the distal tip of glidewire remains in its position.
If the guidewire is not exchange length, remove the wire from the shuttle sheath, introduce it into the H1 catheter, and then advance this system into the sheath.
Using fluoroscopy, continue to advance the slip catheter until it emerges from the distal tip of the shuttle sheath.
Ensure that during navigation the glidewire is in the lead.
Navigate the glidewire into the common carotid artery (CCA) and advance the slip catheter over it.
Use roadmapping, as necessary.
Ensure that the manipulation remains proximal to the stenosed segment.
If greater wire purchase is needed, advance the glidewire into the external carotid artery not internal carotid artery (ICA).
Once the slip catheter is in the proximal CCA, advance the shuttle sheath over the slip catheter and glidewire.
After the shuttle sheath is positioned in the CCA, retract and remove the glidewire and slip catheter from the shuttle sheath.
Intervention
Placement of Embolic Protection Device
Perform pre-intervention cervical and cerebral angiography in AP and lateral views.
Select the working views.
Measure the diameter of stenosis and the length of the affected segment.
Measure the vessel diameter proximal and distal to the stenosis.
Use these measurements to select the appropriate size embolic protection device. See Table 10.3.
Table 10.3
Accunet
Filter size fully expanded (mm)
Reference vessel diameter minimum to maximum range (mm)
4.5
3.25–4.0
5.5
4.0–5.0
6.5
5.0–6.0
7.5
6.0–7.0
As an example, if the stenosis is measured as 3.28 mm, use the 4.5-mm embolic protection device.
Preparation of Embolic Protection Device
Transfer the embolic protection system, e.g., RX ACCUNET™ (Guidant) system from its package in an aseptic fashion.
Hold the dispenser hoop containing the embolic protection device and the flush tool in one hand.
Loosen the RHV on the flush tool.
Hold the flush tool upright and attach it to a 10-cc syringe containing heparinized saline (see Fig. 10.1).
Fig. 10.1
Courtesy Abbott™ (Abbott Vascular, Santa Clara, CA)
Flush gently to remove air from the filter and confirm that the fluid exits from the RHV.
After confirming that no part of the light blue distal section of the sheath is within the RHV, tighten the RHV on the delivery sheath (see Fig. 10.1).
Release the flushing tool and delivery sheath from the dispenser hooks.
Also release the proximal end of the wire from the dispenser hook.
Ensure the system is undamaged, prior to its use.
Flush the sheath with heparinized saline to remove air from it. Confirm that the fluid exits from the proximal end of the sheath.
Place a torque device on the proximal end of the guidewire of the embolic protection system and tighten it.
Carefully pull back on the torque device while observing the flushing tool, to pull the filter basket into the delivery sheath, until the distal tip of the delivery sheath aligns with the proximal end of the blue filter obturator (Fig. 10.2). DO NOT pull in the filter basket any further.
Fig. 10.2
Courtesy Abbott™ (Abbott Vascular, Santa Clara, CA)
Loosen the RHV of the flushing tool and slide the flushing tool off the delivery system.
Loosen the torque device and position it so that the proximal end of the light blue delivery sheath is within the central collet tube of the peel away adapter of the torque device. After this positioning, tighten the torque device.
Delivery of Embolic Protection Device
Ensure the guide catheter (shuttle sheath) is well positioned and stable in the CCA.
The tip of the guide catheter must always be kept in view during the procedure to ensure that it does not prolapse into the aortic arch.
If needed, shape the tip of the RX Accunet™ embolic protection system (EPS) guidewire.
Insert the tip of the guidewire into the introducer tool and then advance these as a unit into the RHV of guide catheter, until the delivery sheath has entered the guide catheter.Stay updated, free articles. Join our Telegram channel
Full access? Get Clinical Tree
