Monitoring ENT Procedures

Fig. 16.1
The thyroid gland and surrounding structures

Thyroidectomy

Removal of the thyroid gland is indicated for a variety of conditions including tumor, goiter, or hyperthyroidism among others. Depending on the pathology, one (hemithyroidectomy) or both (total thyroidectomy) lobes may be removed. To access the thyroid and/or parathyroid glands, a horizontal incision is made across the front of the neck followed by a longitudinal incision through the strap muscles. Division of the local vasculature follows division of the musculature so that the lobes of the thyroid can be mobilized. RLN identification is the first goal of IOM for thyroidectomy. Identification of the RLN is essential for the purposes of avoiding the nerve during the procedure. Identification of the RLN is only assured when electrical stimulation results in a recorded compound muscle action potential (CMAP) from the vocalis muscle on the EMG recording.

The Recurrent Laryngeal Nerve

The RLN is a branch of the vagus nerve (CNX) that supplies motor function and sensation to the larynx. It innervates all of the intrinsic muscles of the larynx except the cricothyroid muscle, which is innervated by the superior laryngeal nerve (SLN). The RLN branches from the vagus nerve at the level of the subclavian artery on the right and the aortic arch on the left. After looping under the respective artery, the RLN ascends along the tracheoesophageal groove. The paired nerves are named “recurrent” because after branching, they turn back or run in a direction opposite to the vagus nerve. A minority of patients have a nonrecurrent laryngeal nerve branching off the vagus nerve at the level of the cricoid. During surgery, the RLN can be injured in a number of ways: complete or partial transection, traction, compression, misplaced ligature, thermal injury, or ischemia [3]. If the RLN is injured, it can result in temporary or permanent nerve paralysis. If the damage is unilateral, the patient may wake up with hoarseness. If there is bilateral nerve palsy, the airway may be compromised, resulting in dyspnea and in severe cases the need for a tracheostomy. The RLN also provides sensory innervation to the glottis, and a deficit may result in problems swallowing. Rates of injury range from 1 to 8 %, with significantly increased risk to the RLN when surgery is for reexploration, thyroid carcinoma, and total thyroidectomy [35]. In addition to thyroid and parathyroid procedures, the RLN is often monitored during ACDF, aortic arch procedures, carotid endarterectomy, and posterior fossa surgeries [68]. The SLN can also be injured during surgery [9, 10]. To monitor the SLN, the surgeon must place electrodes/needles in the cricothyroid muscle [9, 10]. Damage to the SLN results in a monotone voice or inability to change pitch.

RLN Monitoring

Spontaneous and triggered EMG recorded from the vocalis muscle is used to monitor the RLN. During thyroid and parathyroid procedures, it is imperative that two channels are available to monitor both the left and right vocal cords. Endotracheal (ET) tubes with left and right electrodes integrated directly into the tube are commercially available. Alternatively, adhesive paired electrodes can be attached to standard ET tubes. Proper placement of the electrodes of the endotracheal tube is of critical importance [2, 11, 12] (Fig. 16.2). Early communication with the anesthesia team will greatly aid in confirming proper electrode placement. The neuromonitorist should request a short acting paralytic for intubation such as succinylcholine as well as avoidance of lidocaine as these drugs will impair early recording ability for the purposes of confirming correct electrode placement [2]. Since the anesthesiologist will be placing the endotracheal tube with electrodes, the monitorist should be able to guide him in proper placement. It should be emphasized to the anesthesiologist that visual confirmation of the electrodes in contact with the vocal cords is essential. The electrode recording surface is often a blue strip or ring depending on the electrodes used. Common misplacements include electrodes that are too superficial or deep as well as a rotated tube [2].
A310119_1_En_16_Fig2_HTML.gif
Fig. 16.2
Proper placement of endotracheal tube electrodes
Unlike EMG from spinal nerve myotomes, baseline EMG recorded from the vocal cords should not be quiet. At baseline, there should be some activity (25–50 μV) present in the recording. This is because the vocal cords are contracted at rest and relax when speaking. Failure to record baseline activity may be due to a number of factors including misplaced tube, use of lidocaine, or residual neuromuscular blockade. Asymmetric baseline activity may indicate that the ET tube and electrodes are rotated and not in contact with one side of the vocal cords. Inadequate baseline recordings will prevent proper monitoring during the case and could result in false-negative results. It is important to correct electrode placement if necessary. This is accomplished by asking the anesthesiologist to move the endotracheal tube while the monitorist views the live EMG recording. As the electrodes move into proper position, the amplitude of recorded activity on the screen will increase. The position showing maximal EMG activity should be marked and the tube secured. It is not unusual for baseline activity to decrease in amplitude during the procedure as a result of changes in electrode impedance resulting from increased secretions.
Identification of the RLN is one of the first steps in the thyroidectomy procedure. One technique, called sweeping, is used to aid the surgeon in initial dissection. Monopolar stimulation is used for the sweeping technique. The surgeon is given a handheld monopolar probe and the monitorist will stimulate continuously (at approximately 2 Hz) while the surgeon sweeps the field in search of the nerve. The presence of a CMAP response indicates that nerve is in proximity. There are technical considerations that the monitorist should be aware of during sweeping. The presence of blood or irrigation in the surgical field may shunt current away from the nerve and prevent a response from being seen despite proximity of the nerve to the stimulator [2]. It is therefore important that the surgical field remain dry when stimulating.
The stimulation parameters for sweeping are different than for direct nerve stimulation. The first difference is the use of a monopolar stimulator versus a bipolar stimulator used for direct nerve stimulation. Monopolar stimulation induces a larger current field and is said to be more sensitive than bipolar stimulation. Bipolar stimulation (having the cathode and anode in close proximity) has a smaller current field and while less sensitive is more specific. The stimulation intensity used for sweeping is higher than for direct nerve stimulation. Continuous stimulation up to 3 mA (pulse width not to exceed 50–100 μs) is performed until a response is seen. Once a response is recorded at supramaximal intensity, the intensity is reduced and the threshold for response determined. If there is no current shunting, the stimulation threshold can be used as an indicator of the distance to the nerve. The response should likewise increase in amplitude as the stimulator approaches the nerve.
When the surgeon is ready to confirm the identity of the RLN, direct nerve stimulation is the optimum method [13, 14]. Direct nerve stimulation uses a bipolar stimulator to find the threshold of activation by increasing the intensity of stimulation in 0.1 mA increments from 0 mA. The pulse width should not exceed 50–100 μs and the stimulation intensity should remain <2 mA. A CMAP recorded from the vocal cords with a latency of approximately 2 ms is confirmation of the identity of the RLN (Fig. 16.3). The stimulation threshold at this point can be used as a comparison to values at closing, possibly offering prognostic information to the surgeon on the function of the RLN.
A310119_1_En_16_Fig3_HTML.gif
Fig. 16.3
Compound muscle action potential in response to stimulation of the RLN. CMAP recorded from the vocal cords bilaterally in response to stimulation of the right RLN. The stimulation intensity was supramaximal at 2.0 mA, causing the large resultant CMAP to be recorded in both channels. Scale bars indicate 1 ms and 100 μV
It is common for the monitorist to not record a response to stimulation even when the surgeon expresses confidence that he is stimulating the RLN. There can be several reasons for this apparent discrepancy. The monitorist should immediately work to confirm that there are no technical issues preventing stimulation and recording. As mentioned earlier, the tube must be properly positioned to insure accurate recording, and this can be confirmed by recording of baseline spontaneous activity on both RLN channels. The presence of a stimulation artifact as well as measurement of current return will serve as confirmation of adequate stimulation. Once technical issues are ruled out, attention should turn to nature of the structure being stimulated. If the surgeon reports seeing a response visually within the field, then he is stimulating a motor nerve or a muscle directly. Often a visual response without EMG confirmation is due to stimulation of the SLN. If this is suspected, a pair of sterile needle electrodes can be handed off to the surgeon and placed in the cricothyroid muscle and an EMG response recorded. If there is no visual evidence of stimulation, then the surgeon may not be stimulating neural tissue or is possibly stimulating a sensory nerve, which will not produce an EMG response.

Only gold members can continue reading. Log In or Register to continue