Recording the Biopotentials of Sleep



Recording the Biopotentials of Sleep


Regina Patrick







The desired signals of brainwaves on an electroencephalogram (EEG), eye movements on an electrooculogram (EOG), heartbeats on an electrocardiogram (ECG), and thoracic and abdominal movements on the respiration channels occur in association with signals that are stronger. The stronger signals may obscure the desired signals. On a polysomnograph (PSG), low-frequency filters and high-frequency filters make it possible to record desired signals while excluding undesired signals. The low-frequency filter setting will not allow any frequency less than that setting to be recorded, and the high-frequency filter setting will not allow any frequency above the setting to be recorded. Thus, any frequency falling between the high- and low-frequency filters will be detected and recorded.

The low- and high-frequency filters are set just below and above, respectively, the desired signal frequency. The American Academy of Sleep Medicine (AASM) (Darien, IL) recommends the following high- and low-frequency filter settings for EEG, EOG, electromyogram (EMG), ECG, and respiratory channels.


RECORDING THE BIOPOTENTIALS OF SLEEP


Electroencephalogram

The EEG frequencies of interest range from less than 1 cycle per second (i.e., <1 Hz) to 14 Hz. Therefore, a low-frequency filter setting of 0.3 Hz (which allows the detection of slow waves) and a high-frequency filter setting of 35 Hz (which allows the detection of alpha waves) are typically used for the EEG channels.


Electrooculogram

A low-frequency filter setting of 0.3 Hz allows the detection of the slow-rolling eye movements that occur with sleep onset, and a high-frequency filter setting of 35 Hz allows the detection of rapid eye movements (REMs) and the fast, low-amplitude EOG activity occurring throughout sleep.


Electromyogram

A low-frequency filter setting of 10 Hz and a high-frequency filter setting of 100 Hz are used to detect EMG frequencies of the skeletal muscles (e.g., legs and chin). This range is sufficient to record the faster firing rate of the muscles (i.e., increased muscle tone) during arousals, wake, or nonrapid eye movement (NREM) sleep and the decreased firing rate of the muscles (i.e., atonia [lack of muscle tone]) during REM sleep.



Electrocardiogram

A low-frequency filter setting of 0.3 Hz and a high-frequency filter setting of 70 Hz are used to detect ECG frequencies. The low-frequency filter setting of 0.3 Hz (i.e., 0.3 cycles per second = 1 cycle per 3.33 seconds) allows the recording of each heartbeat (which normally takes about 0.8 seconds) and a high-frequency filter setting of 70 Hz (i.e., 70 cycles per second = 1 cycle per 0.014 second) allows the recording of a heartbeat’s individual waves (i.e., P, QRS complex, and T wave, the duration of these waves ranges from <0.10 to 0.25 seconds).


Respiratory Effort

The low-frequency filter setting is 0.1 Hz and the high-frequency filter setting is 15 Hz for the respiratory effort sensors, usually, respiratory inductance plethysmography belts. (Some centers may use piezoelectric belts for recording respiratory effort; however, the AASM no longer considers these belts acceptable for recording respiratory effort.) The low setting allows the slower sinusoidal waves of inhalation and exhalation to be recorded, whereas the high setting excludes frequencies that would obscure these waves.


Respiratory Airflow

For the airflow sensors (thermistor or thermocouple, pressure transducer), the low-frequency filter setting is 0.1 Hz and the high-frequency filter setting is 15 Hz. As with the respiratory effort channel, these settings allow the sinusoidal waves of inhalation and exhalation to be recorded while excluding frequencies that would obscure these waves.


PROBLEMS RECORDING BIOPOTENTIAL SIGNALS

Even with the correct filter settings, undesired signals (i.e., artifacts) can appear in a channel and interfere with the recording of a desired signal. The following are some common problems that can lead to artifact in a channel.



Problems Recording EEG Signals


Sweat Artifact

Sweat may cause low-frequency artifact to appear in the EEG, resulting in rolling in the EEG channels. Cooling the patient by increasing air circulation (turning on a fan or air conditioner, opening the door) may eliminate rolling by ameliorating sweating. However, sweat contains salt and the salt that remains on the skin after the skin cools can cause chemical changes in the electrolyte gel used in the electrode. This factor can cause continued problems with a poor signal. If a poor signal continues, the poorly recording electrode(s) should be replaced.


Movement Artifact

Muscle activity during movement can obscure underlying EOG and EEG signals. During movement, it may be necessary for the technologist to describe exactly what is happening—especially if a patient is being monitored for nocturnal seizures or REM sleep behavior disorder (RBD) (i.e., a sleep disorder in which muscle atonia does not occur during REM sleep, thereby allowing the person to act out dreams).


Problems Recording EOG Signals

The retina has a negative charge (reflected by an upward deflection on the EOG channel) and the cornea has a positive charge (reflected by a downward deflection on an EOG channel). The difference between the two, called the “corneoretinal potential,” is approximately -60 mV. The corneoretinal potential makes it possible to record eye movements from sensors placed on the outer canthus of each eye. When the eyes move in the same direction, one electrode will be close to the retina and record a negative charge, whereas the other electrode will be close to the cornea and record a positive charge.

To ensure equal amplitude of conjugate eye movements, the AASM recommends placing one EOG electrode 1 cm above and 1 cm lateral to the midline of the outer canthus of one eye (typically the right eye) and one EOG electrode below and 1 cm lateral to the midline of the outer canthus of the other eye (typically the left eye). In general, both EOG electrodes are referenced
to the same mastoid site (M2); however, it is acceptable for each EOG electrode to be referenced to the contralateral ear mastoid reference (i.e., E1-M2 and E2-M1). If both electrodes are placed above or below the midline of the outer canthus, the EOG signals will deflect in phase (i.e., in the same direction) rather than out of phase (i.e., in opposite directions). When using the recommended placement, the out-of-phase deflections of the EOG signals are greater for horizontal eye movements than for vertical eye movements.






Figure 35-1 The recommended (A) and alternate (B) electrode placement for the electrooculogram channels.

An alternate placement allowed by the AASM for the EOG electrodes is to place each electrode 1 cm below and lateral to the midline of the outer canthus of its respective eye. Both electrodes are referenced to Fpz. In this configuration, vertical eye movements are in phase and horizontal eye movements are out of phase (Figs. 35-1 and 35-2).






Figure 35-2 Eye deflections with the American Academy of Sleep Medicine (AASM)-recommended electrode placement (A) and the AASM alternate electrode placement (B). The deflections with alternate placement are larger for the vertical movements, compared with those of the recommended placement.

An advantage of using the AASM’s recommended eye placement is that all conjugate eye movements produce out-of-phase deflections. However, the deflections of vertical eye movements are greater than those of horizontal eye movements and eye blinks. As a result, it can be difficult to determine the direction of all eye movements, and eye movements with low amplitude may be missed. Using the alternate AASM eye placement, the vertical and horizontal eye movement deflections are
more prominent, compared with those of the recommended placement, and the signals for eye blinking are more readily detected (as in-phase deflections). However, with the alternate placement, the Fpz electrode may allow some EEG activity in the EOG recording, which can make it difficult to distinguish EEG activity from EOG activity, particularly during vertical eye movements. Even if electrodes are correctly placed using the recommended or alternate EOG electrode placement, several physiologic conditions can affect the recording of biopotential changes in an EOG channel.


An Artificial Eye

An artificial eye does not generate the electrical potential changes as would a natural eye. Therefore, an electrode placed by the artificial eye does not reflect eye movements of the affected eye. However, recording eye movements during REM sleep is still possible. Rather than placing an electrode on the outer canthus of each eye, the technologist can place both electrodes on the unaffected eye. The first electrode should be placed 1 cm from the edge and 1 cm above the midline of the outer canthus and the second electrode (which would have been used on the channel of the affected eye) should be placed 1 cm below the midline of the outer canthus of the same eye. Thus, the electrodes will still pick up the deflections needed to record REMs.

If a technologist were to place an electrode by the outer canthus of the affected eye, the EOG signals on that channel would appear dampened or distorted because the electrode continues to record frontal EEG activity. The technologist should make a note in the patient’s chart that the patient has an artificial eye and indicate whether the EOG electrodes are placed on one eye or on both eyes. If the technologist places an electrode by the affected eye, he or she needs to describe how the signal is affected (e.g., “the patient has an artificial right eye; the right EOG signal is dampened”).






Figure 35-3 Electrooculogram (EOG) recording of a patient who suffered a right eye injury. (Image used with permission of Regina Patrick, RPSGT, RST, and obtained on the SomnoStar system, version 9 [2008; Cardinal Health Respiratory Technologies, Palm Springs, CA].)


Eye Diseases That Affect the Retina

Damage to or destruction of the rods or cones in the retina alters these cells’ ability to undergo a photochemical reaction. The retina’s electrical potential consequently decreases, and this factor can affect the corneoretinal potential. A diminished corneoretinal potential may dampen the EOG signal of one or both eyes in people who have diseases that affect the retina, such as Best disease (i.e., congenital macular dystrophy) (Fig. 35-3). The technologist may need to increase the sensitivity of the channels to record the EOG signal. The technologist should make a note in the patient’s chart if sensitivity was increased for the affected eye.


Eye Muscle Problems

Six muscles in the eye, which are arranged as three sets of opposing muscles, allow a person to move the eyes in all four directions (up, down, left, and right) and obliquely. If a problem such as a stroke affects how these muscles move, one or both eyes will not move correctly. As a result of impaired muscle movement, EOG signals may be altered (nonexistent or dampened). The technologist may need to increase the sensitivity of the EOG channels to record the signal. The technologist should make a note in the patient’s chart if sensitivity was increased for the affected eye.



Problems Recording EMG Signals

Problems recording biopotential changes on the chin or leg EMG channels usually result from improper electrode placement, from high impedances (e.g., the skin is insufficiently cleaned), or from electrodes becoming dislodged during sleep. To reduce the likelihood of poor signal quality in the EMG channels, the technologist must show the same care in preparing the electrode sites as when applying EEG and EOG electrodes, and he or she should obtain a low-impedance reading (<10 Ohms [Ω]). The electrodes must also be securely attached to the chin or legs to prevent their being easily dislodged during the night. For example, an extra piece of tape may be placed on the leg sensors to ensure that they remain on the patient during the night. However, vigorous movement by a patient may dislodge an electrode, no matter how well secured. If the PSG recording system provides re-referencing capabilities, the EMG channel can be re-referenced to a backup electrode. If necessary, electrodes may need to be reapplied.


Problems Recording ECG Signals

Problems with the ECG recording are generally caused by poorly attached or dislodged ECG electrodes. If an ECG electrode becomes dislodged during the study (Fig. 35-4), an ECG tracing may be temporarily obtained from alternate electrode derivations, provided that the PSG recording system offers re-referencing capabilities. If re-referencing is not possible, the electrode will need to be reapplied.






Figure 35-4 Poor electrocardiogram (ECG) signal. (Image used with permission of Regina Patrick, RPSGT, RST, and obtained on the SomnoStar system, version 9 [2008; Cardinal Health Respiratory Technologies, Palm Springs, CA].)


Problems Recording Respiratory Effort

People with certain neurologic problems that affect the abdominal and thoracic muscles or who have chest deformities such as kyphosis (commonly called “hunchback”) may be unable to fully expand and contract the chest wall. This insufficient movement may result in reduced respiratory signals, even if the abdominal and thoracic sensors are properly placed on the patient. The technologist may, therefore, need to increase the sensitivity of a respiratory channel to record respiratory signals.


Problems Recording Respiratory Airflow

A common problem with recording airflow pressure changes is that the pressure transducer sensor may become plugged by excessive mucus, especially in a person with an upper respiratory infection or sinus infection. This obstruction will interfere with the proper recording of the signal. The technologist may need to clean or replace the sensor during the study to obtain a clear airflow signal. Another common problem with the airflow transducer sensor is that the air tube becomes kinked or the patient lies on the air tube. The technologist may need to enter the patient’s room to unkink the air tube or dislodge it from under the patient.







Figure 35-5 Eye deflections with the AASM-recommended electrode placement. (Image used with permission of Regina Patrick, RPSGT, RST, and obtained on the SomnoStar system, version 9 [2008; Cardinal Health Respiratory Technologies, Palm Springs, CA].)


PROCEDURES FOR PRE- AND POSTPHYSIOLOGIC CALIBRATIONS

Physiologic calibrations (sometimes called “biocalibrations” or “patient calibrations”) are performed before and after a sleep study. The physiologic calibrations are a series of instructions that are designed to determine that the sensors are correctly recording signals from a patient, the polarity of the signals is correct, the strength of the signal is sufficient for recording, and no unwanted interference exists. The following are features that should appear for each channel during the physiologic calibration procedure.


The EOG Channels

For these channels, it is important that the eye signals of each channel deflect in opposite directions when the person’s eyes move conjunctively (i.e., in the same direction). To mimic the conjunctive eye movements of REM sleep, ask the patient to look left and right five times, then up and down five times (Fig. 35-5). As a result of the corneoretinal potential, the signal on one eye channel should deflect upward and the signal on the other eye channel should deflect downward with each movement. (However, if one uses the alternate EOG electrode placement, horizontal eye movements produce deflections in the opposite direction and vertical eye movements produce deflections in the same direction.)

The next physiologic calibration that involves the eyes elicits alpha waves. To do this, ask the patient to close the eyes for 30 seconds and then open the eyes for 30 seconds. In most people, when the eyes are closed, alpha waves become prominent in the EEG, especially in the occipital channels. When the eyes are opened, the EEG resumes its original waking activity, which consists of a mixture of fast, low-frequency waves, primarily alpha and beta waves (the latter of which have a frequency of 18 to 30 Hz) (Fig. 35-6).

The third physiologic calibration involving the eyes involves having the patient blink the eyes five times rapidly (Fig. 35-7). During wake, a patient may blink at a rate of 0.5 to 2.0 Hz. As drowsiness develops, the blinking rate begins to slow and may be replaced by slow oscillating eye movements.


The Chin EMG Channel

For this channel, it is important to sufficiently detect muscle tone when the person is in NREM sleep. To ensure muscle tone is being recorded appropriately, ask the patient to grit the teeth together or chew (for 5 seconds), and then relax (Fig. 35-8). There should be greater than or equal to 0.5 cm increase in the amplitude of the signal, which reflects the increase in muscle tone. For patients who have no teeth, ask them to open the mouth or yawn instead.


The Snore Channel

For this channel, it is important that the sensor detects the increased vibration of the upper airway muscles that occurs with snoring. Ask the patient to make a snoring noise or hum for 5 seconds (Fig. 35-9). If the patient is unable to make a sufficient snoring noise, an alternative is to ask the patient to make a vocalization such as counting “1, 2, 3,” or saying “A, B, C.” The signal should increase briefly with each vocalization.







Figure 35-6 Electroencephalogram changes with eyes closed (left) and open (right). (Image used with permission of Regina Patrick, RPSGT, RST, and obtained on the SomnoStar system, version 9 [2008; Cardinal Health Respiratory Technologies, Palm Springs, CA].)






Figure 35-7 The appearance of the electrooculogram channels with eye blinking. EEG, electroencephalogram. (Image used with permission of Regina Patrick, RPSGT, RST, and obtained on the SomnoStar system, version 9 [2008; Cardinal Health Respiratory Technologies, Palm Springs, CA].)


The Leg EMG Channels

For these channels, it is important to detect the leg movements that occur with periodic leg movement (PLM) disorder or restless legs syndrome. Some patients with PLM disorder may bend the legs at the knee or hip level; however, more often, a person with PLM disorder may flex the ankle or toes. Therefore, it is important that the leg sensors are able to detect the smallest leg movements (e.g., flexing of the toes) occurring during a sleep study. To mimic PLMs, ask the patient to flex the foot/raise the toes of the right (or left) foot five times and then to flex the foot/raise the toes of the opposite foot
five times (Fig. 35-10). There should be a burst of activity with each flex.






Figure 35-8 Change in chin electromyogram activity with gritting. (Image used with permission of Regina Patrick, RPSGT, RST, and obtained on the SomnoStar system, version 9 [2008; Cardinal Health Respiratory Technologies, Palm Springs, CA].)






Figure 35-9 Simulation of snoring. (Image used with permission of Regina Patrick, RPSGT, RST, and obtained on the SomnoStar system, version 9 [2008; Cardinal Health Respiratory Technologies, Palm Springs, CA].)


The Airflow Channels

For these channels, it is important that the airflow sensor (i.e., thermistor or thermocouple) and pressure transducer sensor are detecting airflow and air pressure changes, respectively, occurring with inhalation and exhalation. It is also important that these sensors detect the lack of airflow during an apnea episode. To ensure that the airflow sensors are detecting nasal breaths and oral breaths, ask the patient to breathe through the nose only for 10 seconds and then to breathe for 10 seconds through the mouth only (Fig. 35-11). As the person inhales and exhales, the thermistor or thermocouple should generate sinusoidal signals on the airflow channel. Nasal pressure transducers typically only measure nasal airflow; therefore, the signal in this channel should be flat during oral breathing maneuvers.

To mimic apnea, ask the patient to take a deep breath in and hold it for 10 seconds. The airflow and air pressure signals should be flat because no air is passing through the nose or mouth (Fig. 35-12).







Figure 35-10 Change in leg electromyogram activity with toe flexes. (Image used with permission of Regina Patrick, RPSGT, RST, and obtained on the SomnoStar system, version 9 [2008; Cardinal Health Respiratory Technologies, Palm Springs, CA].)






Figure 35-11 Oral/nasal airflow tests. (Image used with permission of Regina Patrick, RPSGT, RST, and obtained on the SomnoStar system, version 9 [2008; Cardinal Health Respiratory Technologies, Palm Springs, CA].)







Figure 35-12 Simulation of apnea. (Image used with permission of Regina Patrick, RPSGT, RST, and obtained on the SomnoStar system, version 9 [2008; Cardinal Health Respiratory Technologies, Palm Springs, CA].)


The Abdominal and Thoracic Effort Channels

For these channels, it is important that the sensors detect the movements of the abdomen and chest during normal respiration and during an apnea. During the airflow tests, the abdominal and thoracic effort channels should have a sinusoidal signal moving in correlation with the airflow signal for inhalation and exhalation. When the patient is holding the breath to mimic apnea, the abdominal and thoracic channel signals should be flat (see Fig. 35-12).

Another respiratory feature usually tested during physiologic calibrations is paradoxical breathing—the movement of the abdomen and thorax in opposite directions—when the upper airway is fully blocked but the person continues to make respiratory efforts to breathe. Paradoxical breathing can occur during an obstructive sleep apnea (OSA) episode. To mimic paradoxical breathing, ask the person to hold the breath while making breathing movements. An alternative is to ask the person to pant. Either of these actions will cause the thoracic and abdominal channel signals to move in opposite directions (Fig. 35-13).


The ECG Signal

The ECG signal is not tested as part of the physiologic calibrations procedure. However, the technologist should note that the ECG signal has the correct polarity. The P wave should deflect upward, the QRS complex should have a down-up-down deflection, and the T wave should deflect upward. If the waves are in the wrong directions, the polarity should be corrected before lights out (Figs. 35-14 and 35-15).


ARTIFACT RECOGNITION AND RESPONSE DURING POLYSOMNOGRAPHY

The biggest contributor to artifact on a PSG during a study is patient movement. Patient movement increases muscle activity to the point that the activity obscures the EEG and EOG signals as artifact (Fig. 35-16).

Only gold members can continue reading. Log In or Register to continue

Stay updated, free articles. Join our Telegram channel

Dec 12, 2019 | Posted by in NEUROLOGY | Comments Off on Recording the Biopotentials of Sleep

Full access? Get Clinical Tree

Get Clinical Tree app for offline access