Technical Considerations for MIBG Cardiac Scintigraphy



Fig. 14.1
Structures of gamma camera and the effect of high-energy photons penetration between low-energy collimator and medium-energy collimator on the results of heart-to-mediastinum ratio. 123I emits not only 159 keV photons but also high-energy photons of more than 400 keV [~2.87 %, main contributor 529 keV (1.28 %)], which have a potential to penetrate the thinner collimator septa resulting scatter detected in the 159 keV energy windows. Thus, high-energy photons penetration will more profoundly affect the heart-to-mediastinum (H/M) ratio assessed by low-energy (LE) collimator, which has thinner septa than that by medium-energy (ME) collimator. Although the use of ME provides lower spatial resolution than LE, H/M ratios are assessed from the counts in relatively large regions compared to brain imaging. Thus, semiquantitative cardiac 123I-MIBG scintigraphy can be best performed using ME collimators.



The collimator, which is an important part that determines the quality of SPECT imaging, has continued to be used since Anger first developed a gamma camera in the 1950s. The most common collimator used in nuclear medicine is a parallel-hole collimator, which has developed into a myriad of holes aligned in parallel with each other with a thin lead plate (height number cm) as a partition wall. The shape of the holes, e.g., hexagonal, round, and square, is determined by the production method of the collimator.

Collimators are classified into two major groups: low energy (LE) and medium energy (ME). The thickness of the partition wall (septa) of collimators is designed so that the ratio of the gamma rays passing through it is less than 5 %. Therefore, ME collimators have a thicker partition wall than LE collimators [5] to prevent the penetration of higher-energy gamma rays. The increased septal thickness of ME collimators contributes to increase image resolution and decrease scatter in the images compared to LE collimators but also degrades count sensitivity and spatial resolution. Moreover, nuclear laboratories are not always equipped with ME collimators, but use LE collimators instead because they are suitable for technetium (99mTc), which is used most commonly in clinical practice.

Actually, in addition to LE and ME collimators, various types of collimators are also available, depending on the purpose of the study, in order to achieve good balance among sensitivity, resolution, and appropriate energy extent: low-energy high resolution (LEHR), low-energy high sensitivity, low-energy general purpose (LEGP), medium-energy general purpose, high-energy general purpose, and high-energy pinhole collimators [17, 19]. With respect to 123I-MIBG cardiac scintigraphy, most clinical and published studies have used LE collimators, such as LEGP or LEHR collimators.

Iodine-123 predominantly emits energy of 159 keV (97 %), which is an optimal imaging energy for an LE collimator, but also emits high-energy photons of more than 400 keV (approximately 2.9 %), with the main contributing photons at 529 keV. Higher-energy 123I photons, mostly at 529 keV, can penetrate the thinner septal wall and contaminate 159 keV imaging data [5, 12, 20] (Fig. 14.1). Thus, an LE collimator, which has a thinner septal wall, can show a lower H/M ratio than an ME collimator. On the basis of these findings, recent guidelines on MIBG imaging from the European Council of Nuclear Cardiology recommend using an ME collimator, which has a thicker septal wall than an LE collimator, for MIBG imaging to minimize scattered radiation noise from high-energy emissions [5], but this recommendation is not usually followed in clinical practice. In addition, a lot of previous studies have used an LE collimator.

Clinically, it is a considerable problem that differences in the collimator can influence the H/M ratio profoundly. To compensate for the resulting corruption of image quantitation related to LE collimator penetration by higher-energy photons, Chen et al. developed a mathematical technique (iterative reconstruction with deconvolution of septal penetration) that appears to improve the quantitative accuracy of cardiac 123I-MIBG uptake in reference to a phantom standard [21]. However, clinical applications have yet to be determined.

On the contrary, Nakajima et al. have endeavored to standardize the heart-to-mediastinum (H/M) ratio assessed by using planar images. They designed a simple and reproducible phantom standard for planar image acquisition [20, 21]. The phantom consisted of multiple slices of the heart, lung, mediastinum, and liver containing a radioisotope as a single component. Using these two phantoms, four H/M ratios (anterior and posterior views for each) were obtained to calculate the regression equation. A linear regression equation was analyzed using the formula y − 1 = Ki × (x − 1). The first step was to convert the H/M ratio generated by the LE collimator to the reference value based on mathematical theory using the conversion coefficient Ki. The second step was to convert this H/M ratio to a standardized H/M ratio with the conversion coefficient Ks, which was defined as the average K value for a typical ME collimator. They verified the initial standardization efforts in ten centers, suggesting that this phantom method could be applied to calibrate the results from ME collimators and LE collimators [22].

More recently, much larger Japanese multicenter studies including 84 institutions for the measurement of planar H/M ratios using standard nuclear cameras from a variety of vendors and collimators also confirmed and extended these findings [19]. On the basis of phantom studies, a conversion coefficient of 0.88 was determined to integrate H/M ratios from all acquisition conditions. Using the reference H/M ratio and conversion coefficients for the system can convert an H/M ratio under various conditions converted to the standard one, irrespective of the collimator used. They also demonstrated that two H/M ratios from one phantom could be comparable to one from two phantoms because the conversion coefficient showed a significantly high correlation with an R2 of 0.997. Standardization of the H/M ratio for 123I-MIBG among various scinticamera-collimator combinations will be useful for not only clinical practice but also multicenter studies.



14.3.3 ROI Setting


Cardiac MIBG uptake is semiquantitatively obtained by calculating an H/M ratio, after drawing ROIs over the heart and the upper mediastinum above the lung apices, but below the thyroid gland, in the planar anterior view. Average counts per pixel in the myocardium are divided by average counts per pixel in the mediastinum, thus generating the H/M ratio.

At least three methods have been used to acquire an H/M ratio [1]. Firstly, square or rectangular ROIs are drawn in the center of the heart and upper mediastinum, and the count per pixel ratio is calculated [23]. Secondly, an ROI is drawn around the epicardial border and the valve plane, including the left ventricular cavity [24]. Thirdly, an ROI encompassing the myocardium alone, tracing the epicardial and endocardial borders, excluding the valve plane and cavity, has also been used [25].

The H/M ratio is a simple measure and the above three measuring methods appear to give similar results in previous reports when they were performed at a single center. However, it is prone to variation among setters with respect to the position of the ROI [13, 26]. As for the mediastinum, differences in ROI location sensitivity can influence the results; therefore, it is desirable to provide a certain set reference.

Okuda, Nakajima, and colleagues developed novel software for semiautomatically measuring the H/M ratio (standardized method for automatic ROI setting in MIBG, smartMIBG) [27]. In this semiautomatic measuring system, the examiner manually sets a circular contour to the cardiac region by checking and modifying the center point and circular ROI displayed on the screen. The mediastinum ROI is set up automatically by using four steps:


  1. 1.


    To determine the trunk right border of the body, after determining the heart ROI manually, the right border of the liver is determined from the count profile curve of the bottom of the ROI as the position of the right border of the trunk.

     

  2. 2.


    To determine the trunk left border, after the upper boundary corresponding to the apex of the lungs is estimated as three times the vertical cardiac diameter from the lower border consistent with the center point of the cardiac ROI, a chest counter profile curve is created to detect the minimal count point on the mediastinum. Subsequently, it is folded back the distance between this point and the trunk right border and defined as the left border.

     

  3. 3.


    To determine the superior mediastinal edge, the apex of the lung is determined by vertically detecting the minimum value on the upper mediastinum, applying thyroid uptake as a reference. If thyroid uptake is too low to analyze, three times the vertical cardiac diameter is employed.

     

  4. 4.


    Finally, the rectangular mediastinal ROI is placed in the upper 30 % part of the mediastinum, and the horizontal size of the mediastinal ROI is defined as 10 % of body width.

     

The test-retest reliability of the H/M ratio calculated using this automated method is better than that for manual analysis. Interobserver agreement was also good using the semiautomated method. The H/M ratios generated with the method were higher than those obtained manually, because the count point of the mediastinum ROI is minimal. By using the semiautomatic method to set an ROI, it is possible to reduce differences between setters and between facilities.


14.3.4 Washout Rate


Washout of cardiac 123I-MIBG has been shown to be an important measure of cardiac sympathetic innervation. The WR is thought to reflect the turnover of catecholamines, which relates to the degree of sympathetic drive, although it represents several mechanisms (at least vesicular exocytosis and extravesicular diffusion in the nerve terminals) [28], and the mechanisms responsible for this phenomenon have still not been completely explained. Increased sympathetic activity is associated with diminished cardiac 123I-MIBG retention on delayed images, resulting in a greater degree of cardiac 123I-MIBG washout.

Only gold members can continue reading. Log In or Register to continue

Stay updated, free articles. Join our Telegram channel

Aug 25, 2017 | Posted by in NEUROLOGY | Comments Off on Technical Considerations for MIBG Cardiac Scintigraphy

Full access? Get Clinical Tree

Get Clinical Tree app for offline access